PENGARUH VARIASI LAMA PENYIMPANAN REAGEN DI DALAM ALAT KIMIA ANALYSER TERHADAP HASIL PEMERIKSAAN KADAR UREUM DARAH DI RS DKT SLAMET RIYADI SURAKARTA

Sugiyarti, Sugiyarti (2023) PENGARUH VARIASI LAMA PENYIMPANAN REAGEN DI DALAM ALAT KIMIA ANALYSER TERHADAP HASIL PEMERIKSAAN KADAR UREUM DARAH DI RS DKT SLAMET RIYADI SURAKARTA. Skripsi thesis, Universitas Setia Budi Surakarta.

[img] Text
INTISARI ABSTRACT.pdf

Download (428kB)
[img] Text
form publikasi sugiyarti.pdf

Download (351kB)
[img] Text
sugiyarti surat 21 agustus'.pdf

Download (39kB)
[img] Text
COVER - BAB 1.pdf

Download (2MB)
[img] Text
BAB 2.pdf

Download (439kB)
[img] Text
BAB 3.pdf
Restricted to Repository staff only

Download (381kB)
[img] Text
BAB 4.pdf
Restricted to Repository staff only

Download (323kB)
[img] Text
BAB 5.pdf
Restricted to Repository staff only

Download (99kB)
[img] Text
DAFTAR PUSTAKA.pdf

Download (214kB)
[img] Text
LAMPIRAN.pdf
Restricted to Repository staff only

Download (212kB)

Abstract

Sugiyarti. 2023. The Effect of Variations in Reagent Storage Duration in Chemical Analyzer on Blood Urea Level Examination Results at DKT Slamet Riyadi Hospital Surakarta. Bachelor’s degree Program in Medical Laboratory Technology, Health Sciences Faculty, Setia Budi University. The number of people with kidney failure in Indonesia in 2013 is estimated at 499,800 population. Screening with blood laboratory test such as examination of urea levels can help early detection and diagnosis of kidney function disorders. Examination of urea levels requires urea reagent. The reagents used in urea examination have different stability times, depending on where they are stored. This study was conducted to determine the effect of variations in reagent storage duration in the chemical analyzer on the results of examining blood ureum levels. This study was conducted using blood samples obtained from 14 respondents with normal urea levels (10 – 50 mg/dL). Each sample received six blood ureum levels checked. Urea reagents are stored in a chemical analyzer with a tray temperature of 2-12oC. Variations in the storage duration of reagents start from days 1, 5, 10, 15, 20, and 25. Results of this study showed that the control serum examination until the 25th day still met the requirements of accuracy and precision. Based on statistical calculations using one-way ANOVA test, a value of p=0,998 (p >0.05) was obtained, which means that there is no significant difference between the 6 measurement gropus (day 1, 5, 10, 15, 20 and 25). In general, there was a decrease in urea levels that varied from each respondent to the 25th day, but there was no statistically significant difference, so it can be concluded that there was no significant effect between variations in reagent storage duration in the chemical analyzer on the results of examining blood urea levels in the study after 25 days.

Item Type: Thesis (Skripsi)
Uncontrolled Keywords: ureum, kimia analyser, stabilitas reagen, urea, chemical analyzer, reagent stability